Drug-Eluting Balloon Disappoints in Bifurcation Lesions

Key Points:
  • Paclitaxel-eluting balloon fails to improve angiographic, clinical outcomes in bifurcation lesions
  • BMS arm performs better than expected
  • Larger trial called for

By Jason Kahn
Wednesday, May 26, 2010

PARIS, France—In the latest setback for an intriguing technology, a paclitaxel-eluting balloon combined with a bare-metal stent (BMS) failed to improve outcomes vs. a BMS or drug-eluting stent (DES) alone in treating bifurcation lesions, according to results presented during a late-breaking trial session May 25 at EuroPCR 2010.

For the Drug-Eluting Balloon In Bifurcation Trial (DEBIUT), researchers led by Pieter R. Stella, MD, PhD, of University Medical Center Utrecht (Utrecht, the Netherlands), randomized 120 patients with coronary bifurcation lesions to 3 treatment arms, all involving provisional stenting and final kissing balloon inflation:

  • Angioplasty using only a BMS
  • Angioplasty with drug-eluting balloon inflation and a BMS
  • Angioplasty using only a paclitaxel-eluting DES

The stent platform used in each arm was the Liberté (Boston Scientific, Natick, MA). The potential benefits of the Dior paclitaxel-eluting balloon (Eurocor, Bonn, Germany), Dr. Stella noted, include local drug delivery over a short period of time, avoidance of chronic inflammation due to absence of polymers, and avoidance of multiple layers of metal.

In particular, DEBIUT researchers wanted to see whether the drug-eluting balloon could help achieve better side branch patency while reducing the need for side branch stenting and for prolonged dual antiplatelet therapy. In the BMS and drug-eluting balloon arms, only 3 months of dual antiplatelet therapy were prescribed.

In DEBIUT, patients in all 3 arms were well matched in terms of baseline demographics and procedural characteristics including fluoroscopy time and total procedure time. Procedural success was 100% in each arm.

Six-month angiographic follow-up, though, showed no difference among the treatment groups for the primary endpoint of late lumen loss in the main branch or side branch (table 1).

Table 1. Late Lumen Loss at 6-Month Follow-Upa

 

BMS

Drug-Eluting Balloon

DES

Proximal Main Branch, mm

0.64

0.42

0.18

Distal Main Branch, mm

0.1

-0.04

-0.11

Side Branch, mm

0.23

0.11

-0.08

a P = NS for all comparisons.

Nevertheless, Dr. Stella pointed out that the numerical reduction in late loss between the balloon and BMS arms, though not statistically significant, was encouraging, especially in the side branch, where the rate was 50% lower with the Dior balloon. Also, binary restenosis was lower with the paclitaxel-eluting balloon vs. BMS in the proximal main branch (0 vs. 3.0%), the distal main branch (9.4% vs. 18.2%), and the side branch (6.3% vs. 12.1%), though these differences were also not statistically significant.

All other angiographic outcomes were equivalent in each arm in the main and side branches.

In terms of clinical outcomes, there were also no differences among any of the treatment arms at 6 months, although drug-eluting balloon patients showed numerical reductions in MACE and TLR compared with BMS patients (Table 2).

Table 2. Clinical Outcomes at 6-Month Follow-Upa

 

BMS

Drug-Eluting Balloon

DES

Stent Thrombosis

0

0

2.5%

TLR

27.0%

12.5%

10.0%

Death

0

0

0

MACE

27.0%

15.0%

17.5%

a P = NS for all comparisons.

“Due to the unanticipated good results in the [BMS] arm, the primary endpoint was not reached, and because of the lack of power, the 50% reduction in side-branch late loss was not statistically significant,” Dr. Stella said. “However, there are strong trends which show a favorable outcome combining [the drug-eluting balloon] with a BMS in the main branch and using a [drug-eluting balloon] in the side branch especially with regards to late loss and binary restenosis rates.”

He added that limiting dual antiplatelet therapy to 3 months in the BMS plus paclitaxel-eluting-balloon patients seemed safe, with no occurrences of stent thrombosis.

“Certainly, a larger randomized study is warranted between the drug-eluting balloon and a cobalt-chromium stent vs. a DES plus plain old balloon angioplasty,” Dr. Stella said.

The Latest Blow

It has been an uneven road for drug-eluting balloon technology. At the 2009 American Heart Association Scientific Sessions, researchers presented results from the PEPCAD III trial showing that the Sequent Please paclitaxel-eluting balloon combined with a BMS (B. Braun Melsungen AG, Vascular Systems, Berlin, Germany) failed to reach noninferiority against a sirolimus-eluting stent in treating de novo coronary lesions and actually showed worse safety outcomes.

And at EuroPCR 2009, researchers reported interim results from the Piccoleto trial, which showed no improvement in percent diameter stenosis in patients with small coronary vessels treated with the Dior balloon vs. Taxus stents. The trial was halted early when the drug-eluting balloon could not prove noninferiority.

But the news has not been all bad. The year before, at EuroPCR 2008, researchers presented data from the PEPCAD I trial showing that the Sequent Please achieved low rates of restenosis and MACE in a cohort with small-vessel coronary disease.

Despite Outcome, Call for a Larger Trial

Commenting on DEBIUT, Sigmund Silber, MD, of Müller Hospital Munich (Munich, Germany), noted that the study was well conducted, using provisional T stenting as a default with stenting of the side branch in less than 8% of cases. In addition, there were no geographic misses with the drug-eluting balloon.

However, there was a mixture of 2 different generations of the Dior balloon used, “and angiographic follow-up was not as high as I would have liked (91.5%),” he added. “The primary endpoint was not reached, maybe because the late loss of the Liberté stent was lower than expected, maybe due to overstretching of the proximal main branch by the kissing balloon technique.”

Nevertheless, despite the outcome of the trial, “I agree with Pieter Stella that it is time now for a major randomized trial,” Dr. Silber said. “But it should be powered for a primary clinical endpoint with no routine angiographic follow-up, comparing 2 drug-eluting balloons with a bare stent plus a dedicated bifurcation DES, and maybe if we choose a noninferiority design, the chances of getting a positive trial are higher.”

 


Source:
Stella PR. Drug eluting balloons in coronary bifurcations: The Drug Eluting Balloon In BifUrcation Trial (DEBIUT). Presented at: EuroPCR; May 25, 2010; Paris, France.

 

Disclosures:

  • Dr. Stella reports serving on the advisory board of Eurocor.
  • Dr. Silber reports receiving grant support or speakers fees, or serving as a consultant or an advisory board member for Abbott, Biotronik, Boston Scientific, Bristol-Myers Squibb, Cordis, Daiichi-Sankyo, Eli Lilly, Eurocor, Medtronic, Orbus Neich, Pfizer, sanofi-aventis, Terumo, and The Medicines Company.

 

Related Stories:



3 Comments

  • avatar

    Dear Dr Kahn,

    I am Dr Vaquerizo, one of the principal investigators of the Spanish Dior Registry. I think, that if you write about DEB, could be interesting to show to the international community, all the results regarding this new concept of PCI. In my opinion, registries give really important and additional information to randomized trials. The Spanish Dior registry was set up to assess the efficay and safety of Dior balloon in some situations at high risk of restenosis or stent thrombosis, in which previous interventions including DES have shown suboptimal results, these situations included ISR and small vessell disease (<2.5mm), including as small vessell side branches of true bifurcation lesions and ostial bifurcated lesions. Preliminay results have been accepted as abstract presentation around the world: oral abstarct presentation in AHA congress 2009 (Orlando), oral and poster abstract presentation in TCT Asia Congress 2010 (Seoul), 2 oral presentations in EuroPCR 2010 (Paris) and as poster presentation in the next European congress 2010. Moreover information comes from a multicenter registry involving 10 experienced Spanish centers using Dior balloon in a real-world population, the only exclusion criteria was acute STEMI, cardiogenic shock and severe lesion calcification. Currenttly, we report interesting results in small vessel disease (mean reference vessel diameter 1.9±0.3mm), including 89 patients, 38% diabetes, 58% presented as ACS, 30% had 3-vessel disease, 30% true bifurcated lesions or ostial bifurcated lesions. At 6 months (FU completed in 46 patients) we report promising results with single digit TLR (2.4%). We have also completed angio follow-up in 30 patients, with 4/30 (12.5%) binary restenosis and 0.27 late loss. We recognize that we need to completed follow-up, but meanwhile, I think that would be interesting to report these results.

    Sincerely,
    Dr Beatriz Vaquerizo
    Hospital del Mar, Barcelone

  • avatar

    I also disagree with the authors Headlines. The clinical results were best in DEB arm, and a 50% reduction of in segment late loss was found.
    Also the author provided a general overview. That should have included the DIOR DEB registry data as presented by Antonio Serra and Beatriz Vaquerizo on behalf of 10 Spanish centers during AHA 2009 and in two lectures during EuroPCR 2010. In segment late loss of 0.27 a TLR rate of 2.2% and 0% Thrombosis in small vessels at 6 months follow up are excellent, and do not justify negative comments.

    kind regards

    Rembert Pogge von Strandmann

  • avatar

    Dear Jason, I would like to disagree with your headlines.
    Although a statistical significance was not reached due to the just 0.63 in segment late loss in the BMS arm ( vs. 0.43), a more than 50 % reduction of in segment late loss in the sidebranch was found.
    Secondly the clinical follow-up showed best results in the DEB arm - and to be honest this is what really matters to our patients.
    Like in several Pepcath trials there seems to be a ~0.40 late loss in DEB + BMS treatment, however in our study this was combined with a better safety outcome for the patients.
    And finally we have to remember that 80% of DEBIUT patients were treated with the first generation DIOR, where 20 % were treated with gen II, with an improved coating method.
    In conclusion I would state that DEB seems a valid alternative in patients where we would not like to use a DES. Furhter studies for various subgroups like AMI etc are of course needed.

Add Comment

Text Only 2000 character limit

Page 1 of 1


Related Content: 


0