Abiomed Identifies Purge-Control Issue With Impella: FDA

The agency says the company has reported 37 serious injuries and three deaths related to the issue as of August 17.

Abiomed Identifies Purge-Control Issue With Impella: FDA

Abiomed/Johnson & Johnson MedTech has sent a letter to customers highlighting a problem with its Automated Impella Controllers (AICs), the US Food and Drug Administration communicated through an Early Alert on August 26, 2026.

AICs function as the primary user interface for the Impella mechanical circulatory support device. They control and monitor Impella’s catheter as well as provide information about its position across the aortic valve.

As of August 17, the company had reported 37 serious injuries and three deaths associated with this issue, the agency said. The affected AICs were sold under product codes 0042-0000-US, 0042-0010-US, and 0042-0040-US.

The alert relates to "purge cassette recognition issues resulting from failures of the purge flag component within the purge pressure sensor assembly,” according to the FDA announcement.

When this component fails, an AIC exchange is required to solve the problem. The process may delay procedures that haven’t yet started, the agency explains. “In cases involving hemodynamic instability, this delay may result in a prolonged period without mechanical circulatory support.”

When failures occur during active use, support “will continue while purge delivery is stopped and the AIC exchange is planned,” the FDA adds. Although this situation is “urgent,” the brief interruption can be prepared for. 

Abiomed plans to address the issue by replacing the purge flag and retainer components within the sensor, either through annual preventive maintenance or individual outreach by the manufacturer’s service team. Customers can expect to be contacted by the company.

In the meantime, Abiomed’s letter instructs users to take several actions, including:

  • If there is a failure of the purge flag component resulting in an alarm, the first step is to reinsert the purge disc; if that fails to clear the alarm, users should switch to a backup AIC.
  • The AIC housing should be cleaned with mild detergent, because improper cleaning can damage the purge flag’s plastic components.
  • Users should continue to perform routine preventive maintenance on an annual basis.

In an emailed statement, a Johnson & Johnson spokesperson said the company "has issued a voluntary recall of Automated Impella Controllers to address a purge cassette recognition issue identified in 0.2% of cases. Hospital inventory can continue to be used to ensure continuity of care. Patient safety remains our top priority, and existing alarms and monitoring support the safe and effective use of Impella."