AGENT IDE Data Confirm the Value of Lesion Prep Before DCB in ISR
A post hoc analysis of 2-year TLF risk found few independent predictors. Only one was modifiable: percent diameter stenosis after lesion preparation.
For patients whose in-stent restenosis (ISR) is treated with the Agent drug-coated balloon (DCB), “relatively few” factors predict the likelihood of target lesion failure by 2-year follow-up, a post hoc analysis of the AGENT IDE trial shows.
The only modifiable factor they found was in-lesion percent diameter stenosis, which the researchers say points to the importance of optimal lesion preparation.
AGENT IDE was the pivotal trial that underpinned the 2024 US Food and Drug Administration approval of the paclitaxel-coated device (Boston Scientific) as the first DCB dedicated to ISR to hit the US market. Since then, the uptake of Agent into clinical practice has been rapid.
The trial’s main results revealed a TLF rate of 17.9% at 1 year, a number that with further follow-up had reached 32.7% at 3 years. This deep dive into predictors of TLF—defined as ischemia-driven target lesion revascularization, target vessel-related MI, or cardiac death—found the 2-year TLF rate was 27.0%
J. Dawn Abbott, MD (Brown University, Providence, RI), lead author of the new paper, told TCTMD that research in this area is evolving.
“In the bare-metal-stent era, we spent so much time trying to understand why restenosis was occurring, and the list of factors that were associated with it, like patient factors, lesion factors, procedural factors, was quite extensive,” she said. “And then when we moved to drug-eluting stents, that list really became quite abbreviated. Certain things we know were prone to failure, like the long lesions and small vessels, are now [the domain] of drug-coated balloons.”
With the availability of DCB to treat ISR, it’s necessary to look at predictors specific to that scenario, added Abbott.
This question is a clinically relevant one, she specified. “When you’re speaking to a patient about requiring a PCI for in-stent restenosis, even with a good initial outcome, you have to have a discussion with them about surveillance. If one in four are going to potentially end up back in the cath lab in 2 years, that’s a large proportion still.”
The clinical takeaway from their analysis is clear, said Abbott. “The less well the vessel is prepped, the more likely [TLF] is.”
TLF at 2 Years
For this analysis, which was published recently in Circulation: Cardiovascular Interventions, the researchers looked at the 406 patients with ISR who had been randomized to DCB treatment. There were no differences between patients who did versus didn’t have TLF in terms of age, sex, lesion length, minimum lumen diameter, and percent diameter stenosis. Also similar were the clinical procedural success rate, overall procedural time, use of intravascular imaging, and use of ancillary devices.
Using demographic, clinical, and core laboratory angiographic data, the researchers identified three independent predictors of higher 2-year TLF risk:
- Multiple stent layers prior to treatment (HR 1.80; 95% CI 1.23-2.65)
- White race (HR 1.71; 95% CI 1.03-2.85)
- In-lesion percent diameter stenosis after lesion preparation (HR 1.16 per 10% increment; 95% CI 1.03-1.31)
Additionally, having a target lesion located in the LAD was tied to lower 2-year TLF risk (HR 0.56; 95% CI 0.36-0.87).
“While limited in scope, this study is unique in that it leverages data from the first randomized controlled trial of DCB therapy for coronary ISR in the United States, with a large, multicenter design that enhances external validity,” Abbott and colleagues note.
Whether or not a patient has multiple stent layers in the lesion being addressed is “unmodifiable at the time of DCB implantation,” they point out. Still, “its strong association with adverse outcomes should prompt broader consideration,” as well as awareness of the need to optimize the initial PCI to prevent ISR in the first place. In these cases, operators should consider both what the long-term ramifications might be of placing additional stents and a possible role for brachytherapy, the authors suggest.
The association between TLF risk and white race, they write, “may be explained by unmeasured factors, including differential clinical care outside the study protocol, rather than true biological predisposition to neointimal hyperplasia.”
Abbott said the incremental link between TLF and percent diameter stenosis as a continuous variable is a reminder that intravascular imaging is needed for proper lesion prep. In AGENT IDE, such imaging was done in 70% of cases and operators were experienced in its use, she noted, so it’s possible failure rates might be higher in real-world practice.
As for what factors might sway the choice between DES and DCB for ISR, Abbott said that in cases already involving multiple stent layers, adding yet another stent may be impossible. With a single stent layer, both DES and DCB may be options, especially given that some studies have shown greater long-term efficacy with DES compared with DCB, she noted.
In her own practice, Abbott takes a flexible approach to these decisions as they arise. “You don’t have to necessarily decide up front,” she said, adding that DES may be preferred in cases where there is existing stent fracture or late neothrombosis with a high degree of intraluminal calcification.
If the lesion being treated is similar to those studied in AGENT IDE, “I think you can embark on a strategy where you’re planning a DCB and then if you cannot obtain optimal vessel prep, then you just convert . . . over to a stent,” Abbott suggested. “ Sometimes that takes a little more time, and of course everyone is concerned with efficiency, but I think that’s a good strategy.”
Notably, the AGENT IDE trial did not enroll patients with chronic total occlusions or lesions measuring 26 mm or longer.
Future research should explore patterns in more complex lesions and whether TLF predictors differ among various mechanisms of failure, Abbott said. “We need to accumulate data to tell operators the best way to optimize. There will still be some failure, but at least there will be more standardization. As these new therapies are being introduced, there’s a lot of variability. . . . We’ll have to see where the field goes.”
Caitlin E. Cox is Executive Editor of TCTMD and Associate Director, Editorial Content at the Cardiovascular Research Foundation. She produces the…